About: Microwave plasma ion sources for selected ion flow tube mass spectrometry: Optimizing their performance and detection limits for trace gas analysis     Goto   Sponge   NotDistinct   Permalink

An Entity of Type : http://linked.opendata.cz/ontology/domain/vavai/Vysledek, within Data Space : linked.opendata.cz associated with source document(s)

AttributesValues
rdf:type
Description
  • The performance of the ion sources used in selected ion flow tube mass spectrometry, SIFT-MS, instruments is paramount in determining their sensitivities and detection limits for trace gas analysis. The microwave discharge plasma ion source that is currently used for the production of currents of the precursor H3O+, NO+ and O2+ ions for SIFT-MS is described, and the ion chemistry occurring within the plasma and the dissociation of the precursor ions on the helium carrier gas are considered. Thus, it is shown that the most suitable ion source gas composition is a mixture comprising maximal water vapour and minimal air at the lowest total pressure at which the discharge is sustained and stable. It is also shown that the injection energies of the precursor ions into the helium carrier gas must be kept low to minimize collisional dissociation of the ions and thus to minimize the fraction of reactive impurity ions in the carrier gas.
  • The performance of the ion sources used in selected ion flow tube mass spectrometry, SIFT-MS, instruments is paramount in determining their sensitivities and detection limits for trace gas analysis. The microwave discharge plasma ion source that is currently used for the production of currents of the precursor H3O+, NO+ and O2+ ions for SIFT-MS is described, and the ion chemistry occurring within the plasma and the dissociation of the precursor ions on the helium carrier gas are considered. Thus, it is shown that the most suitable ion source gas composition is a mixture comprising maximal water vapour and minimal air at the lowest total pressure at which the discharge is sustained and stable. It is also shown that the injection energies of the precursor ions into the helium carrier gas must be kept low to minimize collisional dissociation of the ions and thus to minimize the fraction of reactive impurity ions in the carrier gas. (en)
  • Výkonnost mikrovlnných plazmatických iontových zdrojů pro hmotnostní spektrometrii v proudové trubici s vybranými ionty, SIFT-MS, je zásadní pro zajištění jejich citlivosti a dosažení lepších detekčních limitů pro analýzu stopových plynů. Práce pojednává o iontové chemii probíhající v plynovém výboji ve zdroji používaném pro přípravu iontů H3O+, NO+ a O2+ pro SIFT-MS. Výsledku ukazují, že optimální je směs obsahující maximum vodní páry a minimální množství vzduchu za nejnižšího možného tlaku, při kterém se udrží stabilní výboj. Výsledky dále ukazují že injekční energie musí být udržena pod určitým limitem, tak aby nedocházelo ke kolizní disociaci iontů. Za těchto podmínek je možno dosáhnou iontových signálů větších než milión pulsů za sekundu. Zastoupení nečistot v iontech prekurzoru bylo při tom menší než 1%. Detekční limit metody SIFT-MS se tak dostává do oblasti zředění jedné molekuly na deset miliard molekul vzduchu. (cs)
Title
  • Microwave plasma ion sources for selected ion flow tube mass spectrometry: Optimizing their performance and detection limits for trace gas analysis
  • Mikrovlnné plazmatické iontové zdroje pro hmotnostní spektrometrii v proudové trubici s vybranými ionty. Optimalizace jejich výkonnosti a detekčních limitů pro analýzu stopových plynů (cs)
  • Microwave plasma ion sources for selected ion flow tube mass spectrometry: Optimizing their performance and detection limits for trace gas analysis (en)
skos:prefLabel
  • Microwave plasma ion sources for selected ion flow tube mass spectrometry: Optimizing their performance and detection limits for trace gas analysis
  • Mikrovlnné plazmatické iontové zdroje pro hmotnostní spektrometrii v proudové trubici s vybranými ionty. Optimalizace jejich výkonnosti a detekčních limitů pro analýzu stopových plynů (cs)
  • Microwave plasma ion sources for selected ion flow tube mass spectrometry: Optimizing their performance and detection limits for trace gas analysis (en)
skos:notation
  • RIV/61388955:_____/07:00088033!RIV08-AV0-61388955
http://linked.open.../vavai/riv/strany
  • 117;124
http://linked.open...avai/riv/aktivita
http://linked.open...avai/riv/aktivity
  • P(GA202/06/0776), Z(AV0Z40400503)
http://linked.open...iv/cisloPeriodika
  • 1-3
http://linked.open...vai/riv/dodaniDat
http://linked.open...aciTvurceVysledku
http://linked.open.../riv/druhVysledku
http://linked.open...iv/duvernostUdaju
http://linked.open...titaPredkladatele
http://linked.open...dnocenehoVysledku
  • 433816
http://linked.open...ai/riv/idVysledku
  • RIV/61388955:_____/07:00088033
http://linked.open...riv/jazykVysledku
http://linked.open.../riv/klicovaSlova
  • microwave plasma ion source; selected ion flow tube mass spectrometry; SIFT-MS; breath analysis (en)
http://linked.open.../riv/klicoveSlovo
http://linked.open...odStatuVydavatele
  • NL - Nizozemsko
http://linked.open...ontrolniKodProRIV
  • [843B41791ADF]
http://linked.open...i/riv/nazevZdroje
  • International Journal of Mass Spectrometry
http://linked.open...in/vavai/riv/obor
http://linked.open...ichTvurcuVysledku
http://linked.open...cetTvurcuVysledku
http://linked.open...vavai/riv/projekt
http://linked.open...UplatneniVysledku
http://linked.open...v/svazekPeriodika
  • 267
http://linked.open...iv/tvurceVysledku
  • Dryahina, Kseniya
  • Smith, D.
  • Španěl, Patrik
http://linked.open...n/vavai/riv/zamer
issn
  • 1387-3806
number of pages
is http://linked.open...avai/riv/vysledek of
Faceted Search & Find service v1.16.118 as of Jun 21 2024


Alternative Linked Data Documents: ODE     Content Formats:   [cxml] [csv]     RDF   [text] [turtle] [ld+json] [rdf+json] [rdf+xml]     ODATA   [atom+xml] [odata+json]     Microdata   [microdata+json] [html]    About   
This material is Open Knowledge   W3C Semantic Web Technology [RDF Data] Valid XHTML + RDFa
OpenLink Virtuoso version 07.20.3240 as of Jun 21 2024, on Linux (x86_64-pc-linux-gnu), Single-Server Edition (126 GB total memory, 47 GB memory in use)
Data on this page belongs to its respective rights holders.
Virtuoso Faceted Browser Copyright © 2009-2024 OpenLink Software