  • A bilateral comparison between the Centro Español de Metrología (CEM) and the Český Metrologický Institut (CMI) of radiance temperature scale realizations in the range from 232°C to 1085°C was carried out during 2012 to support the calibration measurements capabilities of CMI in radiation thermometry. The CEM capabilities were demonstrated previously in a recent comparison of European laboratories over the range from 156°C to 1000°C . A CMI KE-LP5 radiation thermometer, working at 1568 nm, was used as a traveling standard. CEM measurements were done at the fixed-points (FPs) of Zn, Ag, and Cu and, for the rest of the temperatures, variable temperature blackbodies (VTBBs) were used. CMI measurements were done at the FPs of Sn, Al, and Cu, and the rest of the temperatures were measured with VTBBs. The size-of-source effect was measured at CEM to decide whether or not the measurements from both laboratories should be corrected by this effect (when the diameter of the sources was different at each laboratory). CMI performed the measurement of the Al FP before and after CEM to evaluate the stability of the radiation thermometer. The results for both laboratories are summarized, and they agree within their expanded uncertainties.
  • A bilateral comparison between the Centro Español de Metrología (CEM) and the Český Metrologický Institut (CMI) of radiance temperature scale realizations in the range from 232°C to 1085°C was carried out during 2012 to support the calibration measurements capabilities of CMI in radiation thermometry. The CEM capabilities were demonstrated previously in a recent comparison of European laboratories over the range from 156°C to 1000°C . A CMI KE-LP5 radiation thermometer, working at 1568 nm, was used as a traveling standard. CEM measurements were done at the fixed-points (FPs) of Zn, Ag, and Cu and, for the rest of the temperatures, variable temperature blackbodies (VTBBs) were used. CMI measurements were done at the FPs of Sn, Al, and Cu, and the rest of the temperatures were measured with VTBBs. The size-of-source effect was measured at CEM to decide whether or not the measurements from both laboratories should be corrected by this effect (when the diameter of the sources was different at each laboratory). CMI performed the measurement of the Al FP before and after CEM to evaluate the stability of the radiation thermometer. The results for both laboratories are summarized, and they agree within their expanded uncertainties. (en)
Title
  • Bilateral Comparison Between CMI and CEM in Radiance Temperature Scale Realization from TeX to TeX
  • Bilateral Comparison Between CMI and CEM in Radiance Temperature Scale Realization from TeX to TeX (en)
skos:prefLabel
  • Bilateral Comparison Between CMI and CEM in Radiance Temperature Scale Realization from TeX to TeX
  • Bilateral Comparison Between CMI and CEM in Radiance Temperature Scale Realization from TeX to TeX (en)
skos:notation
  • RIV/00177016:_____/14:#0001021!RIV15-MSM-00177016
http://linked.open...avai/riv/aktivita
http://linked.open...avai/riv/aktivity
  • P(7AX12096)
http://linked.open...iv/cisloPeriodika
  • 3-4
http://linked.open...vai/riv/dodaniDat
http://linked.open...aciTvurceVysledku
http://linked.open.../riv/druhVysledku
http://linked.open...iv/duvernostUdaju
http://linked.open...titaPredkladatele
http://linked.open...dnocenehoVysledku
  • 5263
http://linked.open...ai/riv/idVysledku
  • RIV/00177016:_____/14:#0001021
http://linked.open...riv/jazykVysledku
http://linked.open.../riv/klicovaSlova
  • Bilateral comparison; Radiance temperature scale; Standard radiation thermometer (en)
http://linked.open.../riv/klicoveSlovo
http://linked.open...odStatuVydavatele
  • DE - Spolková republika Německo
http://linked.open...ontrolniKodProRIV
  • [E90506C998C0]
http://linked.open...i/riv/nazevZdroje
  • International Jurnal of Thermophysics
http://linked.open...in/vavai/riv/obor
http://linked.open...ichTvurcuVysledku
http://linked.open...cetTvurcuVysledku
http://linked.open...vavai/riv/projekt
http://linked.open...UplatneniVysledku
http://linked.open...v/svazekPeriodika
  • 35
http://linked.open...iv/tvurceVysledku
  • Strnad, Radek
  • Kňazovická, Lenka
  • Campo, D. del
  • Martín, M. J.
issn
  • 1572-9567
number of pages
Faceted Search & Find service v1.16.116 as of Feb 22 2024


Alternative Linked Data Documents: ODE     Content Formats:   [cxml] [csv]     RDF   [text] [turtle] [ld+json] [rdf+json] [rdf+xml]     ODATA   [atom+xml] [odata+json]     Microdata   [microdata+json] [html]    About   
This material is Open Knowledge   W3C Semantic Web Technology [RDF Data] Valid XHTML + RDFa
OpenLink Virtuoso version 07.20.3239 as of Feb 22 2024, on Linux (x86_64-pc-linux-gnu), Single-Server Edition (126 GB total memory, 82 GB memory in use)
Data on this page belongs to its respective rights holders.
Virtuoso Faceted Browser Copyright © 2009-2024 OpenLink Software