About: Experimental setup for the humidity sensors calibration     Goto   Sponge   NotDistinct   Permalink

An Entity of Type : http://linked.opendata.cz/ontology/domain/vavai/Vysledek, within Data Space : linked.opendata.cz associated with source document(s)

AttributesValues
rdf:type
Description
  • The calibration setup was prepared for the various humidity and dew point sensors calibration in the air, nitrogen and carbon dioxide environment. The humidity is adjusted by the climate chamber Vötsch VC 2020 (air measurements) and by the set of bubblers (nitrogen and CO2). The highly precise dew-point meter Dew-Master is used as a reference for the calibration fits calculations. The flow of the gas through the reference sensor is maintained by the additional air circulating system. The gas flow ensures that the humidity doesn't condensate in the on the sensor sensitive element. The signals from sensors that are being calibrated and from the sensors that are monitoring the temperature sensors are communicated over the ELMB IO module into the computer where they are processed by the DAQ system prepared in the PVSS SCADA software. The whole setup is highly mobile (except the climate chamber) and can be used for the calibration of up to 5 sensors in one batch.
  • The calibration setup was prepared for the various humidity and dew point sensors calibration in the air, nitrogen and carbon dioxide environment. The humidity is adjusted by the climate chamber Vötsch VC 2020 (air measurements) and by the set of bubblers (nitrogen and CO2). The highly precise dew-point meter Dew-Master is used as a reference for the calibration fits calculations. The flow of the gas through the reference sensor is maintained by the additional air circulating system. The gas flow ensures that the humidity doesn't condensate in the on the sensor sensitive element. The signals from sensors that are being calibrated and from the sensors that are monitoring the temperature sensors are communicated over the ELMB IO module into the computer where they are processed by the DAQ system prepared in the PVSS SCADA software. The whole setup is highly mobile (except the climate chamber) and can be used for the calibration of up to 5 sensors in one batch. (en)
Title
  • Experimental setup for the humidity sensors calibration
  • Experimental setup for the humidity sensors calibration (en)
skos:prefLabel
  • Experimental setup for the humidity sensors calibration
  • Experimental setup for the humidity sensors calibration (en)
skos:notation
  • RIV/68407700:21220/10:00173291!RIV11-MSM-21220___
http://linked.open...avai/riv/aktivita
http://linked.open...avai/riv/aktivity
  • P(LA08032)
http://linked.open...vai/riv/dodaniDat
http://linked.open...aciTvurceVysledku
http://linked.open.../riv/druhVysledku
http://linked.open...iv/duvernostUdaju
http://linked.open...onomickeParametry
  • nejsou stanoveny
http://linked.open...titaPredkladatele
http://linked.open...dnocenehoVysledku
  • 258328
http://linked.open...ai/riv/idVysledku
  • RIV/68407700:21220/10:00173291
http://linked.open...terniIdentifikace
  • Kalib_hum-sensors
http://linked.open...riv/jazykVysledku
http://linked.open...vai/riv/kategorie
http://linked.open.../riv/klicovaSlova
  • humidity; sensors; calibration; ELMB; SCADA (en)
http://linked.open.../riv/klicoveSlovo
http://linked.open...ontrolniKodProRIV
  • [1317B352C659]
http://linked.open...okalizaceVysledku
  • Laboratoř ústavu a CERN
http://linked.open...in/vavai/riv/obor
http://linked.open...ichTvurcuVysledku
http://linked.open...cetTvurcuVysledku
http://linked.open...vavai/riv/projekt
http://linked.open...UplatneniVysledku
http://linked.open...echnickeParametry
  • Kalibrační komora s DAQ systémem, komora 0-100% rel. vhkosti, roztoky kalibračních solí. Vlastníkem výsledku je ČVUT v Praze, IČ 68407700.
http://linked.open...iv/tvurceVysledku
  • Vacek, Václav
  • Vítek, Michal
  • Doubek, M.
  • Janda, M.
http://linked.open...avai/riv/vlastnik
http://linked.open...itiJinymSubjektem
http://localhost/t...ganizacniJednotka
  • 21220
Faceted Search & Find service v1.16.118 as of Jun 21 2024


Alternative Linked Data Documents: ODE     Content Formats:   [cxml] [csv]     RDF   [text] [turtle] [ld+json] [rdf+json] [rdf+xml]     ODATA   [atom+xml] [odata+json]     Microdata   [microdata+json] [html]    About   
This material is Open Knowledge   W3C Semantic Web Technology [RDF Data] Valid XHTML + RDFa
OpenLink Virtuoso version 07.20.3240 as of Jun 21 2024, on Linux (x86_64-pc-linux-gnu), Single-Server Edition (126 GB total memory, 48 GB memory in use)
Data on this page belongs to its respective rights holders.
Virtuoso Faceted Browser Copyright © 2009-2024 OpenLink Software