Attributes | Values |
---|
rdf:type
| |
Description
| - Device is suitable for the destructions of the ceramic heads of the hip joint endoprosthesis with outer diameter 28 mm and with the conical hole. The aim of the destruction in the device is finding the material parameters of the ceramics material because the sizes of the ceramics heads do not allow to make the normalised speciment for the 4-point bending tests. For carrying out the destruction of a set of ceramic heads a special testing device (jig) has been made in which the heads are subjected to compressive load that acts only on the surface of the head's opening (conical hole). In the course of the test, circumferential strains are measured on the head's external surface. On each head, two gauges were stuck (at 180 degree) which, in addition, provided information on whether the deformations in the head are distributed evenly on the circumference - the data of the gauges served for judging the reliability of the results of experiments. The compression load acting on the head was developed by the te
- Device is suitable for the destructions of the ceramic heads of the hip joint endoprosthesis with outer diameter 28 mm and with the conical hole. The aim of the destruction in the device is finding the material parameters of the ceramics material because the sizes of the ceramics heads do not allow to make the normalised speciment for the 4-point bending tests. For carrying out the destruction of a set of ceramic heads a special testing device (jig) has been made in which the heads are subjected to compressive load that acts only on the surface of the head's opening (conical hole). In the course of the test, circumferential strains are measured on the head's external surface. On each head, two gauges were stuck (at 180 degree) which, in addition, provided information on whether the deformations in the head are distributed evenly on the circumference - the data of the gauges served for judging the reliability of the results of experiments. The compression load acting on the head was developed by the te (en)
- Zařízení je určeno k destrukcím keramických hlavic totální kyčelní endoprotézy s vnějším průměrem 28 mm a kuželovým otvorem. Cílem destrukcí hlavic v zařízení je určení materiálových parametrů keramického materiálu, protože rozměry hlavic neumožnují vyrobit normalizované vzorky na 4-bodový test. K zajištění destrukcí keraických hlavic byl navržen a vyrobeno speciální zařízení (přípravek), ve kterém jsou hlavice zatíženy tlakovým zatítením, které působí pouze na kuželovou plochu jejich otvoru. Tím v hlavici vznikají tahová obvodová napětí, která hlavici destruují. Z tohoto důvodu jsou na každé hlavici nalepeny dva tenzometry (po 180 stupních) na její kulové části v blízkosti otvoru. Takto získané údaje o deformaci hlavice slouží jako verifikátor napjatosti v hlavici v průběhu zatěžování a jeho destrukční hodnota slouží jako parametr na který se naladí výpočtový model. Tlakové zatížení působící na hla (cs)
|
Title
| - Destruction device for the ceramic heads with oil medium.
- Destruction device for the ceramic heads with oil medium. (en)
- Zařízení na destrukce keramických hlavic s médiem olej. (cs)
|
skos:prefLabel
| - Destruction device for the ceramic heads with oil medium.
- Destruction device for the ceramic heads with oil medium. (en)
- Zařízení na destrukce keramických hlavic s médiem olej. (cs)
|
skos:notation
| - RIV/00216305:26210/07:PR22729!RIV08-MSM-26210___
|
http://linked.open...avai/riv/aktivita
| |
http://linked.open...avai/riv/aktivity
| |
http://linked.open...vai/riv/dodaniDat
| |
http://linked.open...aciTvurceVysledku
| |
http://linked.open.../riv/druhVysledku
| |
http://linked.open...iv/duvernostUdaju
| |
http://linked.open...onomickeParametry
| - umožnění získat materiálové parametry keramiky i pro tělesa jejichž rozměry neumožňují použití normalizované zkoušky
|
http://linked.open...titaPredkladatele
| |
http://linked.open...dnocenehoVysledku
| |
http://linked.open...ai/riv/idVysledku
| - RIV/00216305:26210/07:PR22729
|
http://linked.open...terniIdentifikace
| |
http://linked.open...riv/jazykVysledku
| |
http://linked.open...vai/riv/kategorie
| |
http://linked.open.../riv/klicovaSlova
| - Ceramic material parameters, experimental device, oil (en)
|
http://linked.open.../riv/klicoveSlovo
| |
http://linked.open...ontrolniKodProRIV
| |
http://linked.open.../licencniPoplatek
| |
http://linked.open...okalizaceVysledku
| |
http://linked.open...in/vavai/riv/obor
| |
http://linked.open...ichTvurcuVysledku
| |
http://linked.open...cetTvurcuVysledku
| |
http://linked.open...UplatneniVysledku
| |
http://linked.open...echnickeParametry
| - Vstupní parametry - vnější průměr hlavice - 28 mm, maximální destrukční tlak - do 90 MPa Výstupní parametry - hodnoty destrukčních přetvoření a destrukčních tlaků, ze kterých se následně vypočítají materiálové parametry keram
|
http://linked.open...iv/tvurceVysledku
| - Fuis, Vladimír
- Návrat, Tomáš
- Houfek, Martin
|
http://linked.open...avai/riv/vlastnik
| |
http://linked.open...itiJinymSubjektem
| |
http://linked.open...n/vavai/riv/zamer
| |
http://localhost/t...ganizacniJednotka
| |