  • Industrial grade platinum resistance thermometers were calibrated in the temperature range 200 °C – 700 °C. Both wire wound and thin film sensor based thermometers were investigated. The calibrations were performed in a vertical Cs-heat pipe furnace and in a horizontal Na heat pipe furnace. The reference thermometer was a SPRT calibrated at fixed points up to Al point. In addition to calibration, various thermal tests including immersion measurements and thermal cycling tests were performed. The stability of the sensors was determined by monitoring the ice point resistance. Possible contamination of the sensors were determined by measuring the resistance ratio R(30 °C)/R(10 °C) several times during the measurement period. The calibration curves were compared with ICE 60751 and ITS-90 reference functions. Considerable changes were found in all tested thermometers.. The wire wound sensors were more stable than the thin film sensors.
  • Industrial grade platinum resistance thermometers were calibrated in the temperature range 200 °C – 700 °C. Both wire wound and thin film sensor based thermometers were investigated. The calibrations were performed in a vertical Cs-heat pipe furnace and in a horizontal Na heat pipe furnace. The reference thermometer was a SPRT calibrated at fixed points up to Al point. In addition to calibration, various thermal tests including immersion measurements and thermal cycling tests were performed. The stability of the sensors was determined by monitoring the ice point resistance. Possible contamination of the sensors were determined by measuring the resistance ratio R(30 °C)/R(10 °C) several times during the measurement period. The calibration curves were compared with ICE 60751 and ITS-90 reference functions. Considerable changes were found in all tested thermometers.. The wire wound sensors were more stable than the thin film sensors. (en)
Title
  • Calibration of Industrial Platinum Resistance Thermometers up to 700 °C
  • Calibration of Industrial Platinum Resistance Thermometers up to 700 °C (en)
skos:prefLabel
  • Calibration of Industrial Platinum Resistance Thermometers up to 700 °C
  • Calibration of Industrial Platinum Resistance Thermometers up to 700 °C (en)
skos:notation
  • RIV/00177016:_____/13:#0001004!RIV15-MSM-00177016
http://linked.open...avai/riv/aktivita
http://linked.open...avai/riv/aktivity
  • P(7A11082)
http://linked.open...vai/riv/dodaniDat
http://linked.open...aciTvurceVysledku
http://linked.open.../riv/druhVysledku
http://linked.open...iv/duvernostUdaju
http://linked.open...titaPredkladatele
http://linked.open...dnocenehoVysledku
  • 64137
http://linked.open...ai/riv/idVysledku
  • RIV/00177016:_____/13:#0001004
http://linked.open...riv/jazykVysledku
http://linked.open.../riv/klicovaSlova
  • Calibration; Industiral; Platinum (en)
http://linked.open.../riv/klicoveSlovo
http://linked.open...ontrolniKodProRIV
  • [3FE83807F63A]
http://linked.open...v/mistoKonaniAkce
  • Madeira
http://linked.open...i/riv/mistoVydani
  • Madeira
http://linked.open...i/riv/nazevZdroje
  • Symposium on temperature and thermal measurements in industry and science
http://linked.open...in/vavai/riv/obor
http://linked.open...ichTvurcuVysledku
http://linked.open...cetTvurcuVysledku
http://linked.open...vavai/riv/projekt
http://linked.open...UplatneniVysledku
http://linked.open...iv/tvurceVysledku
  • Strnad, Radek
  • Hahtela, O.
  • Heinonen, Martti
  • Kajaste, H.
  • Riski, Kr
http://linked.open...vavai/riv/typAkce
http://linked.open.../riv/zahajeniAkce
number of pages
http://purl.org/ne...btex#hasPublisher
  • Neuveden
https://schema.org/isbn
  • 978-972-8574-15-4
Faceted Search & Find service v1.16.118 as of Jun 21 2024


Alternative Linked Data Documents: ODE     Content Formats:   [cxml] [csv]     RDF   [text] [turtle] [ld+json] [rdf+json] [rdf+xml]     ODATA   [atom+xml] [odata+json]     Microdata   [microdata+json] [html]    About   
This material is Open Knowledge   W3C Semantic Web Technology [RDF Data] Valid XHTML + RDFa
OpenLink Virtuoso version 07.20.3240 as of Jun 21 2024, on Linux (x86_64-pc-linux-gnu), Single-Server Edition (126 GB total memory, 48 GB memory in use)
Data on this page belongs to its respective rights holders.
Virtuoso Faceted Browser Copyright © 2009-2024 OpenLink Software