About: Large Volume Sample Stacking for Rapid and Sensitive Determination of Antidiabetic Drug Metformin in Human Urine and Serum by Capillary Electrophoresis with Contactless Conductivity Detection     Goto   Sponge   Distinct   Permalink

An Entity of Type : http://linked.opendata.cz/ontology/domain/vavai/Vysledek, within Data Space : linked.opendata.cz associated with source document(s)

AttributesValues
rdf:type
Description
  • Two CE methods with contactless conductivity detection have been developed for determining the oral antidiabetic drug metformin in human urine and blood. The determination of metformin is performed on a separation capillary with an effective length of 14 cm, using a maximum voltage of 30 kV and with a small injection of 50-fold diluted urine into the capillary. Under these conditions, the migration time of metformin is 35 s and the LOD is 0.3 µM. Large-volume sample stacking was used to determine low metformin levels in serum. The injection of a sample of serum deproteinized with acetonitrile was 10 times greater compared to the injected amount of urine. This enabled reduction of the LOD to 0.03 µM and the metformin migration time equalled 86 s. The undesirable solvent from sample zone was forced out of the capillary to ensure rapidity and good repeatability of the determination. The RSD values for the migration time are 0.1 % for urine and 0.7 % for serum; RSD for the peak areas equalled 1.4 % for urine and 2.6 % for serum. The developed CE technique was tested on performance of routine analyses of metformin in the urine and serum of patients suffering from type II diabetes mellitus.
  • Two CE methods with contactless conductivity detection have been developed for determining the oral antidiabetic drug metformin in human urine and blood. The determination of metformin is performed on a separation capillary with an effective length of 14 cm, using a maximum voltage of 30 kV and with a small injection of 50-fold diluted urine into the capillary. Under these conditions, the migration time of metformin is 35 s and the LOD is 0.3 µM. Large-volume sample stacking was used to determine low metformin levels in serum. The injection of a sample of serum deproteinized with acetonitrile was 10 times greater compared to the injected amount of urine. This enabled reduction of the LOD to 0.03 µM and the metformin migration time equalled 86 s. The undesirable solvent from sample zone was forced out of the capillary to ensure rapidity and good repeatability of the determination. The RSD values for the migration time are 0.1 % for urine and 0.7 % for serum; RSD for the peak areas equalled 1.4 % for urine and 2.6 % for serum. The developed CE technique was tested on performance of routine analyses of metformin in the urine and serum of patients suffering from type II diabetes mellitus. (en)
Title
  • Large Volume Sample Stacking for Rapid and Sensitive Determination of Antidiabetic Drug Metformin in Human Urine and Serum by Capillary Electrophoresis with Contactless Conductivity Detection
  • Large Volume Sample Stacking for Rapid and Sensitive Determination of Antidiabetic Drug Metformin in Human Urine and Serum by Capillary Electrophoresis with Contactless Conductivity Detection (en)
skos:prefLabel
  • Large Volume Sample Stacking for Rapid and Sensitive Determination of Antidiabetic Drug Metformin in Human Urine and Serum by Capillary Electrophoresis with Contactless Conductivity Detection
  • Large Volume Sample Stacking for Rapid and Sensitive Determination of Antidiabetic Drug Metformin in Human Urine and Serum by Capillary Electrophoresis with Contactless Conductivity Detection (en)
skos:notation
  • RIV/00216208:11120/14:43908436!RIV15-MSM-11120___
http://linked.open...avai/riv/aktivita
http://linked.open...avai/riv/aktivity
  • I
http://linked.open...iv/cisloPeriodika
  • June 6
http://linked.open...vai/riv/dodaniDat
http://linked.open...aciTvurceVysledku
http://linked.open.../riv/druhVysledku
http://linked.open...iv/duvernostUdaju
http://linked.open...titaPredkladatele
http://linked.open...dnocenehoVysledku
  • 25673
http://linked.open...ai/riv/idVysledku
  • RIV/00216208:11120/14:43908436
http://linked.open...riv/jazykVysledku
http://linked.open.../riv/klicovaSlova
  • Rapid separation; Large volume sample stacking; Contactless conductivity detection; Capillary electrophoresis; Antidiabetic drug (en)
http://linked.open.../riv/klicoveSlovo
http://linked.open...odStatuVydavatele
  • NL - Nizozemsko
http://linked.open...ontrolniKodProRIV
  • [30BA5CF5B96F]
http://linked.open...i/riv/nazevZdroje
  • Journal of Chromatography A
http://linked.open...in/vavai/riv/obor
http://linked.open...ichTvurcuVysledku
http://linked.open...cetTvurcuVysledku
http://linked.open...UplatneniVysledku
http://linked.open...v/svazekPeriodika
  • 1345
http://linked.open...iv/tvurceVysledku
  • Tůma, Petr
http://linked.open...ain/vavai/riv/wos
  • 000336709300023
issn
  • 0021-9673
number of pages
http://bibframe.org/vocab/doi
  • 10.1016/j.chroma.2014.04.016
http://localhost/t...ganizacniJednotka
  • 11120
Faceted Search & Find service v1.16.118 as of Jun 21 2024


Alternative Linked Data Documents: ODE     Content Formats:   [cxml] [csv]     RDF   [text] [turtle] [ld+json] [rdf+json] [rdf+xml]     ODATA   [atom+xml] [odata+json]     Microdata   [microdata+json] [html]    About   
This material is Open Knowledge   W3C Semantic Web Technology [RDF Data] Valid XHTML + RDFa
OpenLink Virtuoso version 07.20.3240 as of Jun 21 2024, on Linux (x86_64-pc-linux-gnu), Single-Server Edition (126 GB total memory, 112 GB memory in use)
Data on this page belongs to its respective rights holders.
Virtuoso Faceted Browser Copyright © 2009-2024 OpenLink Software